Z-2232-2012 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

FDA device recall Z-2232-2012 was initiated by Terumo Cardiovascular Systems Corporation on Jul 26, 2012 and is designated Class II. Reason for recall: Terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion… The recall status is terminated (terminated Jun 27, 2014). Affected quantity: 232 units.

Recall details

Product Type
Devices
Report Date
Aug 29, 2012
Initiation Date
Jul 26, 2012
Termination Date
Jun 27, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
232 units

Product description

Interface module for CDl System 100/101 Product Usage: The Interface Module for CDl 500 is the interface between the CDl 500 monitor and the system. The module performs the following functions: Communicates primary pump flow data to the CDl 500 Receives blood parameter data from the CDl 500

Reason for recall

Terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the odor of smoke was as associated with the failures. The malfunctions occurred during set-up, during cardiopulmonary bypass, and during routine testing of the modu

Distribution pattern

Worldwide Distribution - US Nationwide including the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, and WV and the countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Germany, Hong Kong, Indonesia, Japan, Jordan, Malaysia, Singapore, South Africa, South Korea, Taiwan, Thailand, UNITED ARAB EMIRATES (UAE), and Vietnam

Code information

catalog number: 802558 and serial numbers:00008-00013, and 00015-00241.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2232-2012
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Terumo Cardiovascular Systems Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Terumo Cardiovascular Systems Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Terumo Cardiovascular Systems… →

Source: openFDA record for Z-2232-2012. Report a data issue: hello@regdatalab.com.