Z-2244-2021 Class I Terminated

Recalled by Cardinal Health 200, LLC — Mansfield, MA

Recall Details

Product Type
Devices
Report Date
August 25, 2021
Initiation Date
June 16, 2021
Termination Date
January 4, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7197 units

Product Description

Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate. Product Code: 43201

Reason for Recall

The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.

Distribution Pattern

US Nationwide distribution.

Code Information

All lot codes