Z-2245-2018 Class II Terminated

Recalled by Synthes (USA) Products LLC — West Chester, PA

FDA device recall Z-2245-2018 was initiated by Synthes (USA) Products LLC on Mar 5, 2018 and is designated Class II. Reason for recall: Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length. The recall status is terminated (terminated Feb 28, 2020). Affected quantity: 7.

Recall Details

Product Type
Devices
Report Date
Jun 27, 2018
Initiation Date
Mar 5, 2018
Termination Date
Feb 28, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7

Product Description

Part no. 690.375, Graphic Case F/6.5mm & 7.3mm Combined Cannulated Screw Set

Reason for Recall

Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.

Distribution Pattern

Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.

Code Information

FL00860

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2245-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.