Z-2254-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- August 9, 2023
- Initiation Date
- June 27, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 75 units
Product Description
Zimmer Periarticular Locking Plate (ZPLP) Distal Lateral Femoral Plate 6 holes 159 mm Length-Indicated for temporary internal fixation and stabilization of osteotomies and fractures, including comminuted fractures, supracondylar fractures, intra-articular and extra-articular condylar fractures, fractures in osteopenic bone, nonunions, and malunions. Item Number: 00235710106
Reason for Recall
Thread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock.
Distribution Pattern
AR AZ MA ME MO NJ NM NY OR WA Foreign: Brazil China Dominican Republic Korea Netherlamds Taiwan
Code Information
Lot Number/UDI: 64254511 /(01)0089024055940(10)64254511; 64296571/(01)0089024055940(10)64296571; 65379087/(01)0089024055940(10)65379087; 65379093/(01)0089024055940(10)65379093; 65379096/(01)0089024055940(10)65379096