Z-2259-2023 Class II Ongoing

Recalled by Philips North America — Cambridge, MA

FDA device recall Z-2259-2023 was initiated by Philips North America on Apr 25, 2023 and is designated Class II. Reason for recall: Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or… The recall status is ongoing. Affected quantity: 20 units OUS.

Recall details

Product Type
Devices
Report Date
Aug 9, 2023
Initiation Date
Apr 25, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 units OUS

Product description

Incisive CT for Brazil SKD-Computed Tomography X-Ray System Model: 728146

Reason for recall

Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down

Distribution pattern

Nationwide Foreign: China Algeria Argentina Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Czech Republic Ecuador Egypt France Germany Greece India Indonesia Iran Iraq Israel Italy Japan Jordan Korea, Republic of Kosovo Latvia Macedonia Mexico Moldova Netherlands Nigeria Pakistan Panama Philippines Poland Romania Russian Federation Saint Lucia Saint Pierre and Miquelon Saudi Arabia Serbia Slovakia Spain Sweden Taiwan Thailand Tunisia Turkey Turkmenistan United Kingdom Uzbekistan Viet Nam

Code information

UDI-DI: (01)00884838104481(21) Serial Numbers: 540009 540008 540015 540012 540016 540017 540003 540004 540005 540006 540007 540001 540011 540000 540014 540013 540018 540010 540002

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2259-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America →

Source: openFDA record for Z-2259-2023. Report a data issue: hello@regdatalab.com.