Z-2261-2016 Class II Terminated

Recalled by The Metrix Company — Dubuque, IA

FDA device recall Z-2261-2016 was initiated by The Metrix Company on Jun 23, 2016 and is designated Class II. Reason for recall: An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of… The recall status is terminated (terminated Apr 11, 2023). Affected quantity: 18,366 cases.

Recall details

Product Type
Devices
Report Date
Aug 3, 2016
Initiation Date
Jun 23, 2016
Termination Date
Apr 11, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18,366 cases

Product description

Container fill and transfer sets, automatic fill compatible, 250 mL, 500 mL, 1000 mL, 1500 mL, 2000 mL, 3000 mL, or 4000 mL.

Reason for recall

An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.

Distribution pattern

Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom

Code information

All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique five character numeric or alphanumeric sequence. Affected product codes for this product: 58740, 58741, 58742, 58749, 58750, 58753, 58754

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2261-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2261-2016. Report a data issue: hello@regdatalab.com.