Z-2263-2012 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 5, 2012
- Initiation Date
- July 23, 2012
- Termination Date
- August 24, 2012
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 unit
Product Description
Cisco brand HealthPresence, version 2.1; Model/Catalog Number: CHP-ENTPRSYR2.I-K9; Product is manufactured and distributed by Cisco Systems, Inc. San Jose, CA Cisco HealthPresence, version 2 is intended for remote consultation of patients with a healthcare provider. Cisco HealthPresence enables remote collaboration between a physician and a patient by providing: high-definition video and advanced audio; transmission and display of third-party medical device data; integration with third party EMR: scheduling: and other software applications for remote consultations and exams. Cisco Health Presence is not intended for use in emergency situations. Cisco HealthPresence is not intended for use in active patient monitoring or alarming.
Reason for Recall
Software anomaly may result in incorrect values and interpretations. The initial vital signs reading will display accurately on the Cisco Health Presence Vitals tab and the OnePlace tab. However, subsequent vital signs readings will only update on the Cisco HealthPresence Vitals tab and within the OnePlace database, but will not update under the OnePlace tab, and thus, will not provide current vit
Distribution Pattern
Nationwide Distribution including the state of TN.
Code Information
Serial/Code Number: 2.1.0.87