Z-2271-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- August 9, 2023
- Initiation Date
- June 28, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 785 units
Product Description
ACCESSRAIL Platform (Standard Blade) Part Number SB-1000, and (Deep Blades) Part Number C-DB-1000. Component of a chest stabilization system for cardiac and chest surgery.
Reason for Recall
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.
Code Information
UDI-DI: 00607567700215