Z-2275-2026 Class II Ongoing

Recalled by Oculus Technologies Of Mexico, S.A. De C.V. — Zapopan

FDA device recall Z-2275-2026 was initiated by Oculus Technologies Of Mexico, S.A. De C.V. on Apr 27, 2026 and is designated Class II. Reason for recall: The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the… The recall status is ongoing. Affected quantity: 5904 units.

Recall details

Product Type
Devices
Report Date
Jun 10, 2026
Initiation Date
Apr 27, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5904 units

Product description

PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050

Reason for recall

The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.

Distribution pattern

US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.

Code information

UDI:00860009268616 Lot: 25G263

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2275-2026
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Oculus Technologies Of Mexico, S.A. De C.V.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2275-2026. Report a data issue: hello@regdatalab.com.