Z-2279-2019 Class II Terminated

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

Recall Details

Product Type
Devices
Report Date
August 21, 2019
Initiation Date
May 30, 2019
Termination Date
May 5, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
247,000 tubes

Product Description

BD Vacutainer Fluoride Tubes for Blood Alcohol determinations Part Number: 367001 Product Usage: Blood Alcohol determinations

Reason for Recall

Lot has been confirmed to have no additive within the tube. Sample processed without the preservative (additive) in the tube, testing has yielded reliable results if the samples were stored at room temperature for no longer than two days. If the sample was stored for more than 2 days, the result for blood alcohol determination might not be accurate (either falsely low or falsely high)

Distribution Pattern

US Nationwide-Distribution in the states of CA, FL, GA, IL, KS, MO, PA, RI, TN Foreign: Canada

Code Information

Lot Number: 8187663 Exp. Date: 2020/7/31 UDI: (01)30382903670018 (17)200731(10)8187663 (30)0100