Z-2280-2026 Class II Ongoing

Recalled by Fresh Roast Systems Inc — Key Biscayne, FL

Recall Details

Product Type
Devices
Report Date
June 17, 2026
Initiation Date
April 15, 2026
Termination Date
N/A
Voluntary/Mandated
FDA Mandated
Product Quantity
38

Product Description

Fresh Roast Systems ColorTrack. Model Name: ColorTrack. Model Number: BENCH R-100

Reason for Recall

A defect has been identified in the manufacture of the Fresh Roast Systems ColorTrack instrument model number BENCH R-100 which could under very rare specific misuse circumstances potentially allow human access to infrared laser light levels in excess of the safety Classification of the product.

Distribution Pattern

U.S.

Code Information

Only Model Number provide. Model Number: BENCH R-100