Z-2283-2012 Class III Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

FDA device recall Z-2283-2012 was initiated by Siemens Healthcare Diagnostics, Inc. on Jun 7, 2012 and is designated Class III. Reason for recall: Siemens has confirmed that the Stratus(R) CS Acute Care(TM) D-Dimer CalPak calibrator lots 502058002 and 501318002 may exhibit SLOPE or CALCV error messages (instrument errors), which indicate… The recall status is terminated (terminated Jan 16, 2014). Affected quantity: 786 cartons.

Recall details

Product Type
Devices
Report Date
Sep 5, 2012
Initiation Date
Jun 7, 2012
Termination Date
Jan 16, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
786 cartons

Product description

Siemens Stratus(R) CS Acute Care(TM) D-dimer Calibrator (DDM CalPak) lots 502058002 and 501318002 The Stratus(R) CS Acute Care(TM) D-dimer Calibrator (DDMR CalPak), Catalog No. CDDMR-C is an in-vitro diagnostic product intended to be used for the calibration of the Stratus(R) CS Acute Care(TM) D-dimer )DDMR) method

Reason for recall

Siemens has confirmed that the Stratus(R) CS Acute Care(TM) D-Dimer CalPak calibrator lots 502058002 and 501318002 may exhibit SLOPE or CALCV error messages (instrument errors), which indicate calibration failures.

Distribution pattern

Nationwide Distribution - USA including AK, AL, AR, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, WA, WI, WV, and WY.

Code information

DDMR CalPak lots 502058002, exp 2012/11/23 and 501318002, exp 2012/08/10

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2283-2012
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2283-2012. Report a data issue: hello@regdatalab.com.