Z-2283-2013 Class II Terminated
FDA device recall Z-2283-2013 was initiated by Stryker Medical Division Of Stryker Corporation on Aug 19, 2013 and is designated Class II. Reason for recall: The nurse call cable (Part Number: QDF23-0571) on the FL23SE Bed is located right under the opening of the bed head box. If an accessory (i.e. IV pole) is inserted into this slot, the circuit… The recall status is terminated (terminated Jan 29, 2014). Affected quantity: 61 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 2, 2013
- Initiation Date
- Aug 19, 2013
- Termination Date
- Jan 29, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 61 units
Product description
The FL23SE Bed is an AC-powered adjustable hospital bed with four built-in electric DC motors and remote controls that is intended for medical purposes. It can be operated by the patient or the caregiver to adjust the height and surface contour of the bed. The bed includes movable and latchable siderails. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms.
Reason for recall
The nurse call cable (Part Number: QDF23-0571) on the FL23SE Bed is located right under the opening of the bed head box. If an accessory (i.e. IV pole) is inserted into this slot, the circuit directing the nurse call system can be damaged.
Distribution pattern
Worldwide Distribution - USA including MI, WI, VA and internationally to Canada.
Code information
Model Numbers: FL23SE with nurse call option: OL235017 & OL235018 List of Affected Devices: J60267 J60425 J60009 J60001 J60094 J60095 J60096 J60348 J60349 J60350 J60351 J60352 J60353 J60354 J60360 J60361 J60362 J60363 J60364 J60365 J60366 J60367 J60368 J60369 J60370 J60371 J60372 J60373 J60374 J60375 J60376 J60377 J60378 J60379 J60380 J60381 J60382 J60383 J60384 J60385 J60386 J60387 J60388 J60389 J60390 J60391 J60392 J60393 J60394 J60395 J60396 J60397 J60398 J60399 J60400 J60401 J60402 J60403 J60404 J60405 J60512
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2283-2013
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026