Z-2284-2026 Class II Ongoing
FDA device recall Z-2284-2026 was initiated by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. on Dec 7, 2023 and is designated Class II. Reason for recall: During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the… The recall status is ongoing. Affected quantity: 16 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 10, 2026
- Initiation Date
- Dec 7, 2023
- Product Quantity
- 16 units
Product description
Reason for recall
Distribution pattern
US Nationwide distribution.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2284-2026
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.