Z-2285-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 17, 2020
- Initiation Date
- May 4, 2020
- Termination Date
- September 20, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 23,824
Product Description
Dimension Vista Enzymatic Creatinine, Siemens Material Number 10700444, UDI Number 00842768036262 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.
Reason for Recall
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.
Code Information
Lot #s 20027AC, 19336AA, 19273AA, 19218AB, 19192AB, 19189AA, 19175AA, ***Updated 8/4/20*** 20083AA ***Updated 11/2/20*** 20128AB 20202AA