Z-2295-2025 Class II Ongoing

Recalled by Merit Medical Systems, Inc. — South Jordan, UT

Recall Details

Product Type
Devices
Report Date
August 20, 2025
Initiation Date
June 5, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6327

Product Description

Custom Manifold Kit: REF: K09-02239C, K09-02249UP, K09-09110B, K09-10294A, K09-11227A, K09-11456B, K09-12123B, K09-13342AP, K09-13391BP, K09-13548, K09-14000A, K09T-12020D, K09T-12275B

Reason for Recall

High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.

Distribution Pattern

US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.

Code Information

REF/UDI-DI/Lot(Expiration): K09-02239C/00884450204313/H3111528(10/31/2027), H3157165(10/31/2027); K09-02249UP/00884450330043/H3116232(8/26/2026), H3156392(8/31/2026), H3173422(8/31/2026), H3178693(8/31/2026), H3190348(3/16/2028), H3195871(3/31/2028), H3205105(4/7/2028); K09-09110B/00884450279076/H3163891(3/21/2028); K09-10294A/ 00884450210000/H3169190(4/30/2027); K09-11227A/00884450210697/H3156393(2/10/2028), H3157208(2/17/2028); K09-11456B/00884450279717/H3112605(8/31/2026), H3156272(8/31/2026); K09-12123B/00884450741146/H3167049(3/2/2028); K09-13342AP/00884450548714/H3119448(10/31/2027), H3127766(10/31/2027), H3152497(1/25/2028), H3152498(1/25/2028), H3157249(1/31/2028), H3170462(3/7/2028), H3196372(3/21/2028), H3205117(4/5/2028); K09-13391BP/00884450752111/H3128003(10/31/2027), H3152494(7/31/2027), H3170418(10/31/2027), H3173459(10/31/2027), H3184951(3/9/2028), H3190567(8/31/2026); K09-13548/00884450582329/H3156627(2/19/2028) K09-14000A/00884450785102/H3125122(11/30/2027), H3136598(11/30/2027) K09T-12020D/00884450860007/T3147714(4/24/2027), T3163752(8/1/2027) K09T-12275B/00884450640500/T3186397(2/22/2027)