Z-2320-2017 Class II Ongoing

Recalled by Nexstim PLC — Helsinki, N/A

Recall Details

Product Type
Devices
Report Date
June 14, 2017
Initiation Date
March 14, 2017
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21

Product Description

NBS System 4 (sw version 4.0 or higher), Software update to 4.3.3 and NBS System 5 (sw version 5.0 or higher), Software update to 5.1.1. The Nexstim Navigated Brain Stimulation System (NBS System) is indicated for noninvasive mapping of the primary motor cortex of the brain to its cortical gyrus. The NBS System provides information that may be used in the assessment of the primary motor cortex for pre-procedural planning. The NBS System is not intended to be used during a surgical procedure. The NBS System is intended to be used by trained clinical professionals.

Reason for Recall

Software defect: the NBS software may accidentally generate duplicate copies of one or several files.

Distribution Pattern

Worldwide Distribution - US to GA only, Foreign: Europe

Code Information

Serial numbers: NBS10391245N4, NBS11011250N4, NBS11071252N4, NBS11201252N4, NBS11451264N4, NBS11491268N4, NBS12051269N4, NBS12261274N4, NBS12341276N4, NBS12441278N4, NBS13341282N4, NBS13431283N4, NBS13451284N4, NBS13481286N4, NBS14011288N4, NBS14361295N5, NBS14481298N5, NBS14491299N5, NBS14511301N5, NBS15131302N5, Serial numbers: NBS10391245N4, NBS11011250N4, NBS11071252N4, NBS11451264N4, NBS11491268N4, NBS12051269N4, NBS12261274N4, NBS12341276N4, NBS12441278N4, NBS13341282N4, NBS13431283N4, NBS13451284N4, NBS14011288N4, NBS14481298N5, NBS14491299N5, NBS14511301N5, NBS15131302N5, NBS15271305N5, NBS15371307N5, NBS16031313N5 and NBS16221316N5. Additionally in Nexstim's warehouse (i.e. not distributed NBS 11201256N4, NBS 13481286N4, NBS 14361295N5 and NBS 16371318N5)