Z-2326-2015 Class II Terminated

Recalled by Aesculap, Inc. — Center Valley, PA

FDA device recall Z-2326-2015 was initiated by Aesculap, Inc. on Jul 7, 2015 and is designated Class II. Reason for recall: A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after… The recall status is terminated (terminated Apr 17, 2017). Affected quantity: 10.

Recall details

Product Type
Devices
Report Date
Aug 12, 2015
Initiation Date
Jul 7, 2015
Termination Date
Apr 17, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10

Product description

S4C Occiput Torque Wrench F/Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with a defined torque 2.8 Nm for tightening the clamping screw in the screw head/hook head/occipital plate/transverse connector and other connectors.

Reason for recall

A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after reprocessing.

Distribution pattern

Distributed US (nationwide) and to Canada.

Code information

product: FW103R, all Lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2326-2015
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Aesculap, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2326-2015. Report a data issue: hello@regdatalab.com.