Z-2326-2015 Class II Terminated
FDA device recall Z-2326-2015 was initiated by Aesculap, Inc. on Jul 7, 2015 and is designated Class II. Reason for recall: A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after… The recall status is terminated (terminated Apr 17, 2017). Affected quantity: 10.
Recall details
- Product Type
- Devices
- Report Date
- Aug 12, 2015
- Initiation Date
- Jul 7, 2015
- Termination Date
- Apr 17, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10
Product description
Reason for recall
Distribution pattern
Distributed US (nationwide) and to Canada.