Z-2327-2025 Class I Ongoing

Recalled by Boston Scientific Corporation — Saint Paul, MN

FDA device recall Z-2327-2025 was initiated by Boston Scientific Corporation on Jul 24, 2025 and is designated Class I. Reason for recall: Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious… The recall status is ongoing. Affected quantity: 247,702 US; 344,670 OUS.

Recall details

Product Type
Devices
Report Date
Aug 27, 2025
Initiation Date
Jul 24, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
247,702 US; 344,670 OUS

Product description

RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696

Reason for recall

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

Distribution pattern

Worldwide Distribution. US nationwide. International distribution worldwide.

Code information

Model 0657 UDI-DI 00802526560934; Model 0682 UDI-DI 00802526560958; Model 0683 UDI-DI 00802526560965; Model 0685 UDI-DI 00802526560972; Model 0686 UDI-DI 00802526560989; Model 0692 UDI-DI 00802526519017 00802526560996 00802526568664 00802526568688; Model 0693 UDI-DI 00802526518850 00802526519055 00802526537769 00802526561009 00802526568701; Model 0695 UDI-DI 00802526519093 00802526561016; Model 0696 UDI-DI 00802526519987 00802526561207 00802526568831

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2327-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Boston Scientific Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Boston Scientific Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Boston Scientific Corporation →

Source: openFDA record for Z-2327-2025. Report a data issue: hello@regdatalab.com.