Z-2330-2015 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-2330-2015 was initiated by Roche Diagnostics Operations, Inc. on Jun 24, 2015 and is designated Class II. Reason for recall: The affected S2 fluid packs may generate a transponder error when the fluid pack is loaded on to the cobas b 221 system. The error prevents the fluid from being recognized by the system and the… The recall status is terminated (terminated Jan 12, 2016). Affected quantity: 19 affected devices of lot 21450781 were distributed in the US.

Recall Details

Product Type
Devices
Report Date
Aug 12, 2015
Initiation Date
Jun 24, 2015
Termination Date
Jan 12, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
19 affected devices of lot 21450781 were distributed in the US.

Product Description

S2 Fluid Pack on the Cobas b 221 system- Catalog Number 03260925184 Blood Gas and Electrolytes Analyzer; The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, bilirubin, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.

Reason for Recall

The affected S2 fluid packs may generate a transponder error when the fluid pack is loaded on to the cobas b 221 system. The error prevents the fluid from being recognized by the system and the instrument cannot start measurements until the affected S2 Fluid Pack is replaced.

Distribution Pattern

US Distribution: CT, TX, LA, ND, MI, NJ, OH, OK, PR, SC, and KY.

Code Information

Catalog Number: 03260925184; Lot Numbers & Exp. Dates: 1. 21450781, Exp. 08/31/2016 2. 21450887, Exp. 08/31/2016 3. 21451391, Exp. 09/30/2016

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2330-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.