Z-2330-2016 Class II Terminated

Recalled by Dako Denmark A/S — Glostrup

FDA device recall Z-2330-2016 was initiated by Dako Denmark A/S on Jul 6, 2016 and is designated Class II. Reason for recall: The pH of Target Retrieval Solution in the kit measures 4-5 instead of the correct pH of 9.1. The recall status is terminated (terminated Oct 13, 2016). Affected quantity: 147 kits.

Recall details

Product Type
Devices
Report Date
Aug 10, 2016
Initiation Date
Jul 6, 2016
Termination Date
Oct 13, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
147 kits

Product description

EnVision FLEX+, Mouse, High pH (Link), Product No. K8002, K8004. Intended for use in immunohistochemistry together with Autostainer Link Instruments.

Reason for recall

The pH of Target Retrieval Solution in the kit measures 4-5 instead of the correct pH of 9.1.

Distribution pattern

Worldwide Distribution -- US, Canada, and Colombia.

Code information

K8002 Lot no: 10112879 K8004 Lot no: 10111463

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2330-2016
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Dako Denmark A/S? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2330-2016. Report a data issue: hello@regdatalab.com.