Z-2333-2012 Class II Terminated

Recalled by Aesculap Implant Systems LLC — Center Valley, PA

FDA device recall Z-2333-2012 was initiated by Aesculap Implant Systems LLC on Jul 25, 2012 and is designated Class II. Reason for recall: There is a possibility that the screw channel that holds the rod in place may be out of tolerance. When the depth of the channel is not deep enough, the rod may not be able to provide ample… The recall status is terminated (terminated Mar 18, 2014). Affected quantity: 5.

Recall details

Product Type
Devices
Report Date
Sep 19, 2012
Initiation Date
Jul 25, 2012
Termination Date
Mar 18, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product description

Aesculap S4 Element 4.5mm Polyaxial Screws (ST240T, ST241T, ST242T, ST243T, ST244T, and ST245T). The S4 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for treatment of the following acute and chronic instabilities or deformities: 1 - degenerative disc disease, 2 - spondylolisthesis, 3 - trauma, 4 - spinal stenosis, 5 - deformities or curvatures, 6 - tumor, 7 - pseudoarthrosis, and 8 - failed previous fusion.

Reason for recall

There is a possibility that the screw channel that holds the rod in place may be out of tolerance. When the depth of the channel is not deep enough, the rod may not be able to provide ample compression on the insert preventing the screw head from locking down to the bone screw.

Distribution pattern

Nationwide Distribution-including the states of IL, MN, NV, OK, and TX.

Code information

All distributed lots and serial numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2333-2012
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Aesculap Implant Systems LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Aesculap Implant Systems LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Aesculap Implant Systems LLC →

Source: openFDA record for Z-2333-2012. Report a data issue: hello@regdatalab.com.