Z-2334-2012 Class II Terminated

Recalled by JK Products & Services, Inc — Jonesboro, AR

FDA device recall Z-2334-2012 was initiated by JK Products & Services, Inc on Jul 8, 2005 and is designated Class II. Reason for recall: The firm has tracked a systematic failure in the door cord on the Sun|Dash Radius 252 product line. In some instances when the door cord is installed with little or no slack outside the upper decor… The recall status is terminated (terminated Jan 13, 2017). Affected quantity: 1948 units.

Recall details

Product Type
Devices
Report Date
Sep 19, 2012
Initiation Date
Jul 8, 2005
Termination Date
Jan 13, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1948 units

Product description

SUN|DASH RADIUS 252 tanning system; 52 lamps, Ultraviolet Radiation. The device is intended to be used for tanning of the human skin.

Reason for recall

The firm has tracked a systematic failure in the door cord on the Sun|Dash Radius 252 product line. In some instances when the door cord is installed with little or no slack outside the upper decor piece, the cord has been known to fail causing a short in the wiring at the pivot point in the cord.

Distribution pattern

Worldwide Distribution - USA (nationwide) and Canada

Code information

PN 8000809705; Serial numbers 248-1298

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2334-2012
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for JK Products & Services, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for JK Products & Services, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about JK Products & Services, Inc →

Source: openFDA record for Z-2334-2012. Report a data issue: hello@regdatalab.com.