Z-2337-2014 Class I Terminated

Recalled by Customed, Inc — Fajardo, PR

FDA device recall Z-2337-2014 was initiated by Customed, Inc on May 20, 2014 and is designated Class I. Reason for recall: Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to… The recall status is terminated (terminated Aug 14, 2017). Affected quantity: 24 lots; 648 units (multiple units per lot).

Recall details

Product Type
Devices
Report Date
Sep 10, 2014
Initiation Date
May 20, 2014
Termination Date
Aug 14, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
24 lots; 648 units (multiple units per lot)

Product description

LAPAROSCOPY GYN PACK - (1) TABLE COVER REINFORCED 44" X 78" (1) LAPAROSCOPY SHEET "T" SMS (3) TOWELS ABSORBENT 15" X 20" (1) SURGICAL BLADE#11 STAINLESS STEEL (1) MAYO STAND COVER REINFORCED (1) SYRINGE 10cc WITHOUT NEEDLE LUER LOCK (2) LEGGINS WITH 7" CUFF 30" x 42" (1) UNDERBUTTOCK DRAPE WITH POUCH 44" X 35" (4) DRAPE UTILITY WITH TAPE (2) GOWN IMPERVIOUS REINFORCED LARGE SMS (1) CONNECTING TUBING 3/16" X 12" EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

Reason for recall

Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.

Distribution pattern

Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.

Code information

Product code 900-618, 24 lots: 109010109 109030609 109040846 109051047 109061208 109071536 109071688 109092073 109102337 112020491 112030732 112051999 112062499 112083436 112093647 113036805 113057495 113057617 113078338 113088878 113099446 113109861 140211734 140312178

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2337-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Customed, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Customed, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Customed, Inc →

Source: openFDA record for Z-2337-2014. Report a data issue: hello@regdatalab.com.