Z-2343-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 28, 2019
- Initiation Date
- July 9, 2019
- Termination Date
- April 23, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 44
Product Description
MAGNETOM Skyra magnetic resonance diagnostic devices Model # 10432915
Reason for Recall
Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.
Distribution Pattern
AL AR AZ CA CO CT DC DE FL GA GU IA ID IL IN KS KY LA MA MD ME MI MN MO MT NC NE NH NJ NY OH OK OR PA SC TN TX VA WA WI WV WY
Code Information
Serial # 46066 45732 46182 45735 45805 46171 45693 45886 45988 45881 46144 46044 46176 46179 46032 46238 145161 46127 45774 45968 46196 46175 45681 46007 46021 46105 46142 45928 45815 46150 145302 45878 46239 46026 145403 45667 145536 45970 45839 45982 46145 46298 45777 46294