Z-2343-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 1, 2021
- Initiation Date
- July 30, 2021
- Termination Date
- July 25, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 units US; 17 units OUS
Product Description
Ingenia 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781342
Reason for Recall
Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia, Spain, Turkey, Ukraine, United Kingdom.
Code Information
S/N and UDI: 85275 (01)00884838055339(21)85275; 85278 (01)00884838055339(21)85278; 85394 (01)00884838055339(21)85394. OUS: 85393 (01)00884838055339 85332 (01)00884838055339 85334 (01)00884838055339 85336 (01)00884838055339 85349 (01)00884838055339 85350 (01)00884838055339 85355 (01)00884838055339 85370 (01)00884838055339 85371 (01)00884838055339 85382 (01)00884838055339 85384 (01)00884838055339 85385 (01)00884838055339 85398 (01)00884838055339 85348 (01)00884838055339 85149 (01)00884838055339 85367 (01)00884838055339 85388 (01)00884838055339