Z-2353-2026 Class II Ongoing

Recalled by Micromed S.p.A. — Mogliano Veneto, N/A

Recall Details

Product Type
Devices
Report Date
June 17, 2026
Initiation Date
May 4, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
207 units

Product Description

SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02

Reason for Recall

Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of FL, PA.

Code Information

UDI-DI: 8033928120897; Lot Code: Model No SD LTM 64 PLUS; UDI-DI 8033928120897; Serial Numbers BAA-0051/01-19, BAA-0069/02-19, BAA-0070/02-19, BAA-0071/02-19, BAA-0082/02-19, BAA-0083/02-19, BAA-0084/02-19, BAA-0085/02-19, BAA-0086/02-19, BAA-0087/02-19, BAA-0088/02-19, BAA-0089/02-19, BAA-0090/02-19, BAA-0091/02-19, BAA-0092/02-19, BAA-0093/02-19, BAA-0095/02-19, BAA-0096/02-19, BAA-0098/02-19, BAA-0104/02-19, BAA-0105/02-19, BAA-0106/02-19, BAA-0107/02-19, BAA-0108/02-19, BAA-0109/02-19, BAA-0110/02-19, BAA-0111/02-19, BAA-0112/02-19, BAA-0113/02-19, BAA-0115/02-19, BAA-0116/02-19, BAA-0117/02-19, BAA-0118/02-19, BAA-0119/02-19, BAA-0121/02-19, BAA-0122/02-19, BAA-0142/03-19, BAA-0155/03-20, BAA-0156/03-20, BAA-0191/04-20, BAA-0192/04-20, BAA-0193/04-20, BAA-0194/04-20, BAA-0286/04-21, BAA-0287/04-21, BAA-0288/04-21, BAA-0289/04-21, BAA-0329/04-22