Z-2355-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- August 16, 2023
- Initiation Date
- July 3, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 24 units
Product Description
LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 24 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-24/00
Reason for Recall
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
Distribution Pattern
US Nationwide distribution.
Code Information
UDI-DI: 04026575182183 Lot Numbers: B922115 B933170 C005042 C211018 C240049