Z-2359-2019 Class II Terminated

Recalled by Agilent Technologies, Inc. — Santa Clara, CA

Recall Details

Product Type
Devices
Report Date
September 4, 2019
Initiation Date
August 15, 2018
Termination Date
September 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
235

Product Description

MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass.

Reason for Recall

When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.

Distribution Pattern

US Nationwide. distribution.

Code Information

Software versions B.07 or below