Z-2371-2014 Class I Terminated

Recalled by Customed, Inc — Fajardo, PR

FDA device recall Z-2371-2014 was initiated by Customed, Inc on May 20, 2014 and is designated Class I. Reason for recall: Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to… The recall status is terminated (terminated Aug 14, 2017). Affected quantity: 16 lots; 512 units (multiple units per lot).

Recall details

Product Type
Devices
Report Date
Sep 10, 2014
Initiation Date
May 20, 2014
Termination Date
Aug 14, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 lots; 512 units (multiple units per lot)

Product description

ARTHROSCOPY PACK- (1) VIDEO CAMERA DRAPE 13cm X 244cm LIF (2) BANDAGE ELASTIC 6" X 5yrd. LIF (1) U-DRAPE 60" X 70" WITH TAPE SPLIT (1) DURAPREP SURGICAL SOLUTION 26m I LIF (1) STOCKINETIE IMPERVIOUS (1) TUBE SUCTION CONNECT. 3/16" X 12' LIF (1) SHEET ARTHROSCOPY T STD SMS WITH POUCH LIF (1) BAG SUTURE FLORAL LIF (1) BLADE SURGICAL #11 CARBON STEEL (2) GOWN IMPERVIOUS EXTRA REINFORCED X-LARGE LIF (1) SYRINGE 30cc SLIP TIP LIF (1) TABLE COVER 44" x 90" (1) MAYO STAND COVER REINFORCED LIF (4) TOWELS ABSORBENT 15" X 20" LIF (5) LAP SPONGES PRE-WASH XRD UF (1) DRAPE SHEET 41 " x 58" SMS UF (1) NEEDLE HYPODERMIC 22G X 1 Y:z (1) NEEDLE HYPODERMIC 18G X 1 Y:z (1) SYRINGE 10cc WITHOUT NEEDLE LUER LOCK LIF (10) GAUZE SPONGES 4" X 4" 16PL Y EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

Reason for recall

Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.

Distribution pattern

Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.

Code information

Product code 900-1062, 16 lots: 109071569 112072765 112083154 113057380 113057684 113057854 113078233 113078670 113089154 113099586 131110151 131110593 131210956 140412452 140513063 140513390

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2371-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Customed, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Customed, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Customed, Inc →

Source: openFDA record for Z-2371-2014. Report a data issue: hello@regdatalab.com.