Z-2373-2016 Class II Terminated

Recalled by Ivoclar Vivadent, Inc. — Amherst, NY

Recall Details

Product Type
Devices
Report Date
August 17, 2016
Initiation Date
July 11, 2016
Termination Date
September 22, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US: 86 units total

Product Description

OptraPol Polishing Discs, Rx ONLY: OptraPol Refill Cup /10: REF/Product ID 634459, OptraPol Refill Small Flame/10: REF/Product ID 634458, OptraPol Assortment: REF/Product ID 634464 --Device Listing # B106952. One step polishing system for all common dental composites and amalgam.

Reason for Recall

After a few seconds of polishing with moderate force, parts of the OptraPol cup are breaking away. OptraPol Small Flame is also affected. Two complaints were filed referring to OptraPol, claiming that the abrasive polishers wear faster than expected.

Distribution Pattern

Distributed in the states of CA, FL, GA, IA, IL, IN, KS, KY, MA, MI, NV, NY, OH, PA, TX, and WA.

Code Information

REF/Product ID 634459: Lot UL0843; REF/Product ID 634458: Lot VL0704; REF/Product ID 634464: Lot VL0705.