FDA device recall Z-2383-2015 was initiated by Philips Electronics North America Corporation on Mar 23, 2015 and is designated Class II. Reason for recall: The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the… The recall status is terminated (terminated Jan 17, 2017). Affected quantity: 62.

Recall details

Product Type
Devices
Report Date
Sep 2, 2015
Initiation Date
Mar 23, 2015
Termination Date
Jan 17, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
62

Product description

Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The DigitalDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctors or technicians.

Reason for recall

The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the Microsoft Windows 7 operating system.

Distribution pattern

US Nationwide Distribution to the states of : Arizona, Minnesota, State of Washington, Wyoming and Ohio. Internationally to Canada.

Code information

DigitalDiagnost

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2383-2015
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2383-2015. Report a data issue: hello@regdatalab.com.