Z-2383-2015 Class II Terminated
FDA device recall Z-2383-2015 was initiated by Philips Electronics North America Corporation on Mar 23, 2015 and is designated Class II. Reason for recall: The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the… The recall status is terminated (terminated Jan 17, 2017). Affected quantity: 62.
Recall details
- Product Type
- Devices
- Report Date
- Sep 2, 2015
- Initiation Date
- Mar 23, 2015
- Termination Date
- Jan 17, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 62
Product description
Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The DigitalDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctors or technicians.
Reason for recall
The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the Microsoft Windows 7 operating system.
Distribution pattern
US Nationwide Distribution to the states of : Arizona, Minnesota, State of Washington, Wyoming and Ohio. Internationally to Canada.