Z-2389-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 8, 2021
- Initiation Date
- July 23, 2021
- Termination Date
- July 26, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4270
Product Description
Marco Shoulder Stabilization Kit, Product Code 711100, (previous product code 100601), Sample Kit S711100
Reason for Recall
Affected products were manufactured with materials found to have a bioburden level exceeding the limit at which the product's sterilization process was validated. As a result, the firm cannot provide full assurance that the affected products were effectively sterilized, and the product may compromise a sterile field and increase the risk of a surgical site infection.
Distribution Pattern
US nationwide distribution. No foreign distribution.
Code Information
All lots manufactured between July 9, 2019 to July 9, 2021