Z-2392-2023 Class II Ongoing

Recalled by Canon Medical System, USA, INC. — Tustin, CA

Recall Details

Product Type
Devices
Report Date
August 16, 2023
Initiation Date
June 22, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12

Product Description

INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE

Reason for Recall

The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.

Distribution Pattern

US Nationwide Distribution: NY, CO, IL, CA, MD, MN, WA

Code Information

UDI-DI: 04987670100192. INFX-8000C, Serial Numbers: BHA1992001, BHB2132003, SWC2242002, AFB1832003, SUC2262008, SUC2292010, BFB1972004, SUB2172002, SWC2262003