Z-2403-2016 Class I Terminated

Recalled by OraSure Technologies, Inc. — Bethlehem, PA

Recall Details

Product Type
Devices
Report Date
August 24, 2016
Initiation Date
July 20, 2016
Termination Date
May 10, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,075 units

Product Description

OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.

Reason for Recall

Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.

Distribution Pattern

Distributed to one customer only in US state of Georgia.

Code Information

Product Code 3001-2807 and the lot number is 6648965. Expiration date October 31, 2016.