Z-2404-2023 Class II Ongoing

Recalled by Philips Medical Systems Nederland B.V. — Eindhoven, N/A

FDA device recall Z-2404-2023 was initiated by Philips Medical Systems Nederland B.V. on Jul 19, 2023 and is designated Class II. Reason for recall: A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation The recall status is ongoing. Affected quantity: 5432 systems in total.

Recall details

Product Type
Devices
Report Date
Sep 13, 2023
Initiation Date
Jul 19, 2023
Voluntary/Mandated
FDA Mandated
Product Quantity
5432 systems in total

Product description

MultiDiagnost-Eleva

Reason for recall

A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation

Distribution pattern

US Nationwide Distribution

Code information

MultiDiagnost-Eleva

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2404-2023
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Medical Systems Nederland B.V.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Medical Systems Nederland B.V.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Medical Systems Nederland B.V. →

Source: openFDA record for Z-2404-2023. Report a data issue: hello@regdatalab.com.