Z-2408-2024 Class II Ongoing
FDA device recall Z-2408-2024 was initiated by Canadian Hospital Specialties Ltd. on Jun 11, 2024 and is designated Class II. Reason for recall: A customer found product with incomplete seals of P/N 20400161S, lot #185198 which can lead to non-sterile product and patient safety concerns. CHS has validated this complaint. The recall status is ongoing. Affected quantity: 6300 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 24, 2024
- Initiation Date
- Jun 11, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6300 units
Product description
Product Name: HIGH FLOW INSUFFLATION TUBING (with Filter) Model/Catalog Number: 20400161S Product Description: Laparoscopic Insufflator
Reason for recall
A customer found product with incomplete seals of P/N 20400161S, lot #185198 which can lead to non-sterile product and patient safety concerns. CHS has validated this complaint.
Distribution pattern
US Nationwide distribution in the state of MA.
Code information
Product code: 20400161S; UDI/DI: 00628725004358 (Pouch), 50628725004353 (Case of 10); Lot: 185198; Expiry: 2028-10-0.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2408-2024
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026