Z-2409-2020 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 8, 2020
- Initiation Date
- May 20, 2020
- Termination Date
- January 7, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2123 total
Product Description
Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 SPANISH, Product Code IAP-0400E
Reason for Recall
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
Distribution Pattern
Worldwide distribution.
Code Information
Serial Number Ranges: 150301V 160119V 160429V through 160430V 160634V 160643V