Z-2412-2020 Class I Terminated

Recalled by Arrow International Inc — Reading, PA

Recall Details

Product Type
Devices
Report Date
July 8, 2020
Initiation Date
May 20, 2020
Termination Date
January 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2123 total

Product Description

Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE ITALIAN, Product Code IAP-0500I

Reason for Recall

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

Distribution Pattern

Worldwide distribution.

Code Information

Serial Number Ranges: AUTOCAT2 WAVE ITALIAN IAP-0500I 160516W 160619W 160804W 160828W 170209W through 170210W 170302W 180309W