Z-2415-2025 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

FDA device recall Z-2415-2025 was initiated by Baxter Healthcare Corporation on Jul 14, 2025 and is designated Class II. Reason for recall: Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the… The recall status is ongoing. Affected quantity: 6 units.

Recall details

Product Type
Devices
Report Date
Sep 3, 2025
Initiation Date
Jul 14, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product description

Spectrum IQ Infusion Pump, Product Code 3570009

Reason for recall

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

Distribution pattern

US distribution to states of: AZ, FL

Code information

UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271