Z-2416-2024 Class II Ongoing

Recalled by EVIDENT SCIENTIFIC INC — Waltham, MA

Recall Details

Product Type
Devices
Report Date
August 7, 2024
Initiation Date
July 22, 2024
Termination Date
N/A
Voluntary/Mandated
FDA Mandated
Product Quantity
N/A

Product Description

SlideView VS200 slide viewer,

Reason for Recall

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

Distribution Pattern

US

Code Information

65159 67105 67104 69153 67106 69154