Z-2417-2020 Class I Terminated

Recalled by Arrow International Inc — Reading, PA

FDA device recall Z-2417-2020 was initiated by Arrow International Inc on May 20, 2020 and is designated Class I. Reason for recall: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in… The recall status is terminated (terminated Jan 7, 2023). Affected quantity: 2123 total.

Recall details

Product Type
Devices
Report Date
Jul 8, 2020
Initiation Date
May 20, 2020
Termination Date
Jan 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2123 total

Product description

Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701

Reason for recall

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

Distribution pattern

Worldwide distribution.

Code information

Serial Number Ranges: AC3 Optimus IABP NA/AJLA IAP-0701 170831F through 170840F 171011F through 171020F 171121F through 171125F 180106F through 180110F 180126F through 180130F 180426F through 180430F 180606F through 180610F 180731F through 180740F 180826F through 180830F 180901F through 180910F 180912F through 180921F 181001F through 181010F 190136F through 190140F 190336F through 190340F 190646F through 190650F 190731F through 190736F 190931F through 190936F 191019F through 191024F 191055F through 191060F 200107F through 200112F 200308F through 200311F

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2417-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Arrow International Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Arrow International Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Arrow International Inc →

Source: openFDA record for Z-2417-2020. Report a data issue: hello@regdatalab.com.