Z-2422-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 26, 2015
- Initiation Date
- August 7, 2015
- Termination Date
- March 1, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 100 units
Product Description
IBC Suction Wand Catalog Number 1990S Product Usage: The IBC Suction Wand is a component of Blood Recovery System for use in open heart procedures, autologous blood recovery and blood suction. It is a disposable, single use component.
Reason for Recall
It was discovered that the heat seal placed onto sterile pouches may not meet strength acceptance criteria.
Distribution Pattern
Worldwide - United States Nationwide and the countries of: Germany, Norway, Turkey, Chile, Pakistan, and Republic of South Africa
Code Information
033115-3445