Z-2428-2012 Class II Terminated

Recalled by Integra LifeSciences Corporation — Plainsboro, NJ

FDA device recall Z-2428-2012 was initiated by Integra LifeSciences Corporation on Aug 9, 2012 and is designated Class II. Reason for recall: T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened… The recall status is terminated (terminated Mar 26, 2014). Affected quantity: 175 units.

Recall details

Product Type
Devices
Report Date
Sep 26, 2012
Initiation Date
Aug 9, 2012
Termination Date
Mar 26, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
175 units

Product description

Integra CRW Precision Arc (CRWPRECISE), CRW Arc System (CRWASL), HRAIM Head Ring Assembly with Intubation Mounts The CRW stereotactic system is used in conjunction with the HRAIM head ring is a target-centered system that can be configured to be CT-only or CTIMR compatible. The CRW stereotactic system is a multi-purpose system used for localizing intercranial targets for precisely directing instruments such as: Biopsy forceps RF lesioning electrodes Deep brain electrodes Recording and stimulating electrodes

Reason for recall

T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened, the T-handle could stop in a vertical position that prevents the CRWPRECISE and certain models of the CRW-ASL from seating properly on the head ring although it appears to be fully seated

Distribution pattern

Worldwide distribution: USA (Nationwide) and countries of: Argentina, Australia, Belgium, Canada, China, Czec, Denmark, France, Hong Kong, Hungary, Italy, Japan, Malaysia, Philippines, Singapore, Spain, South Africa, and UK.

Code information

All lot numbers from 2002 to current

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2428-2012
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2428-2012. Report a data issue: hello@regdatalab.com.