Z-2428-2012 Class II Terminated
FDA device recall Z-2428-2012 was initiated by Integra LifeSciences Corporation on Aug 9, 2012 and is designated Class II. Reason for recall: T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened… The recall status is terminated (terminated Mar 26, 2014). Affected quantity: 175 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 26, 2012
- Initiation Date
- Aug 9, 2012
- Termination Date
- Mar 26, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 175 units
Product description
Integra CRW Precision Arc (CRWPRECISE), CRW Arc System (CRWASL), HRAIM Head Ring Assembly with Intubation Mounts The CRW stereotactic system is used in conjunction with the HRAIM head ring is a target-centered system that can be configured to be CT-only or CTIMR compatible. The CRW stereotactic system is a multi-purpose system used for localizing intercranial targets for precisely directing instruments such as: Biopsy forceps RF lesioning electrodes Deep brain electrodes Recording and stimulating electrodes
Reason for recall
T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened, the T-handle could stop in a vertical position that prevents the CRWPRECISE and certain models of the CRW-ASL from seating properly on the head ring although it appears to be fully seated
Distribution pattern
Worldwide distribution: USA (Nationwide) and countries of: Argentina, Australia, Belgium, Canada, China, Czec, Denmark, France, Hong Kong, Hungary, Italy, Japan, Malaysia, Philippines, Singapore, Spain, South Africa, and UK.
Code information
All lot numbers from 2002 to current