Z-2429-2026 Class II Ongoing
FDA device recall Z-2429-2026 was initiated by Medline Industries, LP on Apr 27, 2026 and is designated Class II. Reason for recall: Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application. The recall status is ongoing. Affected quantity: 49,654 kits total.
Recall details
- Product Type
- Devices
- Report Date
- Jun 24, 2026
- Initiation Date
- Apr 27, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 49,654 kits total
Product description
Medline convenience kit: (82)LABOR & DELIVERY C-SECTION DYNJ907895F
Reason for recall
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Distribution pattern
US Nationwide distribution. OUS distribution pending.
Code information
UDI-DI 10198459366390 Lot 25KBL873