Z-2431-2023 Class I Ongoing

Recalled by Datascope Corp. — Mahwah, NJ

Recall Details

Product Type
Devices
Report Date
September 6, 2023
Initiation Date
July 31, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9175 units

Product Description

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65

Reason for Recall

Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully inserted back into the cart the battery(ies) will not charge.

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

All Lot Numbers. Model: 0998-00-0800-31, UDI: 10607567109053; Model: 0998-UC-0800-31, UDI: N/A; Model: 0998-00-0800-32,UDI: 10607567111117; Model: 0998-00-0800-33, UDI: 10607567109008; Model: 0998-UC-0800-33, UDI: N/A; Model: 0998-00-0800-34, UDI: 10607567111940; Model: 0998-00-0800-35, UDI: 10607567109107; Model: 0998-00-0800-45, UDI: 10607567108421;Model: 0998-00-0800-52, UDI: 10607567108438; Model: 0998-UC-0800-52, UDI: N/A; Model: 0998-00-0800-53, UDI:10607567108391; Model: 0998-UC-0800-53, UDI: N/A; Model: 0998-00-0800-55, UDI: 10607567108414; Model: 0998-UC-0800-55,UDI: N/A; Model: 0998-00-0800-65, UDI: 10607567113432; Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449