Z-2432-2020 Class II Terminated
FDA device recall Z-2432-2020 was initiated by Siemens Healthcare Diagnostics Inc on May 27, 2020 and is designated Class II. Reason for recall: The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes. The recall status is terminated (terminated Jan 12, 2021). Affected quantity: 436.
Recall details
- Product Type
- Devices
- Report Date
- Jul 1, 2020
- Initiation Date
- May 27, 2020
- Termination Date
- Jan 12, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 436
Product description
Reason for recall
Distribution pattern
Foreign distribution only.