Z-2434-2023 Class I Ongoing

Recalled by Datascope Corp. — Mahwah, NJ

Recall Details

Product Type
Devices
Report Date
September 6, 2023
Initiation Date
July 31, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
47 units (42 US, 5 OUS)

Product Description

Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-UC-0800-83, 0998-UC-0800-85,

Reason for Recall

Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

All Lot Numbers. Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449, Model: 0998-UC-0800-75, UDI: N/A; Model: 0998-UC-0800-83, UDI: N/A; Model: 0998-UC-0800-85, UDI: N/A;