Z-2437-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 8, 2020
- Initiation Date
- May 15, 2020
- Termination Date
- May 18, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 186 units
Product Description
(1) CentraLink Data Management System software version 16.0.2 Siemens Material Number (SMN): 11313246 (2) CentraLink Data Management System software version 16.0.3 Siemens Material Number (SMN): 11314337 Product Usage: are clinical laboratory information modules which consolidate test orders and test results, provide test ordering and priority routing information to automation, apply rules to test orders and auto-verification rules for test results, and provide integrated management of quality control of test system.
Reason for Recall
Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review
Distribution Pattern
Worldwide distribution - US Nationwide distribution and the countries of Argentina, Australia, Bahamas, Belgium, Brazil, Chile, Colombia, Czech Republic, Egypt, Germany, Greece, Hungary, India, Italy, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Oman, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
Code Information
Software versions: CentraLink Data Management System software version 16.0.2 UDI: 00630414220635 CentraLink Data Management System software version 16.0.3 UDI: 00630414237268