Z-2437-2025 Class II Ongoing

Recalled by VANTIVE US HEALTHCARE LLC — Deerfield, IL

Recall Details

Product Type
Devices
Report Date
September 3, 2025
Initiation Date
July 28, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4772 units

Product Description

PrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701

Reason for Recall

Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles

Distribution Pattern

Worldwide

Code Information

UDI 00085412639499, All Serial Numbers