Z-2437-2025 Class II Ongoing

Recalled by VANTIVE US HEALTHCARE LLC — Deerfield, IL

FDA device recall Z-2437-2025 was initiated by VANTIVE US HEALTHCARE LLC on Jul 28, 2025 and is designated Class II. Reason for recall: Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles The recall status is ongoing. Affected quantity: 4772 units.

Recall details

Product Type
Devices
Report Date
Sep 3, 2025
Initiation Date
Jul 28, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4772 units

Product description

PrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701

Reason for recall

Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles

Distribution pattern

Worldwide

Code information

UDI 00085412639499, All Serial Numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2437-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for VANTIVE US HEALTHCARE LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2437-2025. Report a data issue: hello@regdatalab.com.